Background and Aims: In recent years high-resolution microendoscopy (HRME) has shown potential to improve screening for esophageal squamous cell neoplasia. Furthering its utility in a clinical setting, especially in lower-resource settings, could be accomplished by reducing the size and cost of the system as well as incorporating the ability of real-time, objective feedback. This article describes a tablet-interfaced HRME with fully automated, real-time image analysis.
Methods: The performance of the tablet-interfaced HRME was assessed by acquiring images from the oral mucosa in a normal volunteer. An automated, real-time analysis algorithm was developed and evaluated using training, test, and validation images from a previous in vivo study of 177 patients referred for screening or surveillance endoscopy in China. The algorithm was then implemented in a tablet HRME that was used to obtain and analyze images from esophageal tissue in 3 patients. Images were displayed alongside the probability that the imaged region was neoplastic.
Results: The tablet-interfaced HRME demonstrated comparable imaging performance at a lower cost compared with first-generation laptop-interfaced HRME systems. In a post-hoc quantitative analysis, the algorithm identified neoplasia with a sensitivity and specificity of 95% and 91%, respectively, in the validation set compared with 84% and 95% achieved in the original study.
Conclusions: The tablet-based HRME is a low-cost tool that provides quantitative diagnostic information to the endoscopist in real time. This could be especially beneficial in lower-resource settings for operators with less experience interpreting HRME images.